X-Series Logger Storage Requirements


The following practices should be carried out when storing an X-Series data logger for an extended period of time:

  • Disconnect the solar tower cable, SP-series solar power pack, or UW6-PW AC power adapter cable from the center 6-pin port on the data logger.
    • Cap the 6-pin port with the blank plug included during shipment.
  • Disconnect any telemetry antenna(s) to prevent any accidental damage resulting from field removal, transport, or storage.
  • Protect all cable and port connections.
    • Inspect all sensor ports and cable connections ensuring there are no signs of moisture or damage.
    • Cover all logger or cable connectors with a dust cap or UW-plug.
    • Keep all system cable O-rings and connectors clean by covering them with the included red (8-pin) or yellow (6-pin) polymer caps. If these caps are no longer accessible, any means of shielding the connectors from debris is acceptable.
    • Use the O-ring grease included in the maintenance kit as needed to re-grease the cable/plug connector O-rings.
  • Store the data logger, SP solar pack or AC adapter, and other logger accessories in a dry, climate-controlled environment.
  • Check the battery voltage on any SP-series solar power packs monthly.
    • Ensure the voltage holds at or above 12V.
    • Place in the sun as needed (normally every 2-3 months) to keep the SLA battery inside topped off.