It may be necessary to reset an iSIC or SDL by manually cycling power. This will allow the system to reboot and can solve various problems. iSIC 1. Open the lid of the iSIC enclosure. 2. Remove both black fuses located to the left of the digital terminal strip. While the fuses are removed, check […]
5100-iSIC
Input/Output Control
The iSIC is capable of adjusting various settings based on a parameters reading. The enable line can be triggered, the sample interval can be changed, and various other configurations adjustments can be made. 1. Open iChart with the current project file. Note: Interrogate all data off of the device as this process will require the […]
Change COM Port
If a different computer than the one used to create the project is being used to download data, it is likely that the COM port will need to be switched. To determine which COM port the data logger is using, please follow the instructions in Verifying iSIC/SDL COM port. 1. Open iChart with the current project file. 2. Right-click […]
Update Firmware With RS-485 Cable
In order to update a SDL’s firmware using a UW-USB-485P communicaiton cable, the communication protocol must be changed to RS-485. 1. Establish temporary RS-485 communicaiton with the device by following the steps in the Direct connection with an SDL500 via RS-485 article. 2. From the menu bar select Edit | Preferences. 3. Switch to the GUI tab. 4. Check the “Enable […]
Changing the Address of an iSIC or SDL500 Data Logger
The address of an iSIC or SDL500 data logger can be changed using iChart software. 1. Open iChart with a current project if not already open 2. From the menu select Advanced | iSIC | iSIC 3. Select the correct PC COM port, Connection type and iSIC Address If the address is unknown and the connection […]
Resetting an Ethernet Password on 5100-iSIC
It is possible to reset an ethernet password on a 5100-iSIC. Note: DeviceInstaller needs to be installed in order to reset the password. It can be installed from the gridconnect CD included with the 5100-iSIC. 1. Power on the 5100-iSIC and connect it to the network. 2. Click Start | Lantronix | DeviceInstaller to open DeviceInstaller. The DeviceInstaller […]
5100-iSIC Quick Start Guide
Overview The NexSens 5100-iSIC (Intelligent Sensor Interface and Control) uses an integrated Ethernet modem to provide a dedicated port for continuous Ethernet-to-router communication. What’s Included Data logger with Ethernet modem 12 VDC battery iSIC Kit (2) Gland fittings (5) Desiccant packs RS-232 interface cable (4) 56 Ω resistors CAT5E Patch cable DB-9 Null modem adapter […]
Changing an iSIC Battery
The battery that powers an iSIC data logger is user-replaceable. 1. Disconnect the fastons from the battery terminals inside the iSIC. 2. Remove the two screws on the battery plate with a #3 Phillips head screwdriver. Store the battery plate and screws. 3. Remove the old battery and put the new battery in its […]
Millivolt (mV) Readings on an Analog Sensor
Analog sensors require a millivolt (mV) reading in order to determine an accurate output. To find these readings, follow the process below. 1. Open iChart without a project file. 2. Select Advanced|iSIC|iSIC from the menu bar. 3. Select the appropriate communication under PC Settings. 4. Type the address of the data logger and click Connect. 5. Click the Analog tab. […]
Connecting an Analog Sensor to the iSIC
In order to connect a sensor to the iSIC, the output voltage must be cut down to 2.5V. In order to accomplish this, two resistors must be utilized to act as a voltage divider. The diagram below demonstrates the proper connection for a 5V sensor. Note: ADx is any single ended iSIC analog inputs […]