Cycling Power to an iSIC or SDL

    It may be necessary to reset an iSIC or SDL by manually cycling power. This will allow the system to reboot and can solve various problems. iSIC 1. Open the lid of the iSIC enclosure. 2. Remove both black fuses located to the left of the digital terminal strip. While the fuses are removed, check […]

    Read More

    Input/Output Control

    The iSIC is capable of adjusting various settings based on a parameters reading. The enable line can be triggered, the sample interval can be changed, and various other configurations adjustments can be made. 1. Open iChart with the current project file. Note: Interrogate all data off of the device as this process will require the […]

    Read More

    Change COM Port

    If a different computer than the one used to create the project is being used to download data, it is likely that the COM port will need to be switched. To determine which COM port the data logger is using, please follow the instructions in Verifying iSIC/SDL COM port. 1. Open iChart with the current project file. 2. Right-click […]

    Read More

    Update Firmware With RS-485 Cable

    In order to update a SDL’s firmware using a UW-USB-485P communicaiton cable, the communication protocol must be changed to RS-485. 1. Establish temporary RS-485 communicaiton with the device by following the steps in the Direct connection with an SDL500 via RS-485 article. 2. From the menu bar select Edit | Preferences. 3. Switch to the GUI tab. 4. Check the “Enable […]

    Read More

    Changing the Address of an iSIC or SDL500 Data Logger

    The address of an iSIC or SDL500 data logger can be changed using iChart software. 1. Open iChart with a current project if not already open 2. From the menu select Advanced | iSIC | iSIC 3. Select the correct PC COM port, Connection type and iSIC Address If the address is unknown and the connection […]

    Read More

    Resetting an Ethernet Password on 5100-iSIC

    It is possible to reset an ethernet password on a 5100-iSIC. Note: DeviceInstaller needs to be installed in order to reset the password. It can be installed from the gridconnect CD included with the 5100-iSIC. 1. Power on the 5100-iSIC and connect it to the network. 2. Click Start | Lantronix | DeviceInstaller to open DeviceInstaller. The DeviceInstaller […]

    Read More

    5100-iSIC Quick Start Guide

    Overview The NexSens 5100-iSIC (Intelligent Sensor Interface and Control) uses an integrated Ethernet modem to provide a dedicated port for continuous Ethernet-to-router communication. What’s Included Data logger with Ethernet modem 12 VDC battery iSIC Kit (2) Gland fittings (5) Desiccant packs RS-232 interface cable (4) 56 Ω resistors CAT5E Patch cable DB-9 Null modem adapter […]

    Read More

    Changing an iSIC Battery

    The battery that powers an iSIC data logger is user-replaceable. 1. Disconnect the fastons from the battery terminals inside the iSIC. 2. Remove the two screws on the battery plate with a #3 Phillips head screwdriver. Store the battery plate and screws.   3. Remove the old battery and put the new battery in its […]

    Read More

    Millivolt (mV) Readings on an Analog Sensor

    Analog sensors require a millivolt (mV) reading in order to determine an accurate output. To find these readings, follow the process below. 1. Open iChart without a project file. 2. Select Advanced|iSIC|iSIC from the menu bar. 3. Select the appropriate communication under PC Settings. 4. Type the address of the data logger and click Connect.   5. Click the Analog tab. […]

    Read More

    Connecting an Analog Sensor to the iSIC

    In order to connect a sensor to the iSIC, the output voltage must be cut down to 2.5V. In order to accomplish this, two resistors must be utilized to act as a voltage divider. The diagram below demonstrates the proper connection for a 5V sensor.   Note: ADx is any single ended iSIC analog inputs […]

    Read More