iChart includes advanced tools to view, diagnose and check sensor and data logger functionality. To access these tools, select Advanced | iSIC | iSIC from the iChart menu system. The iSIC Setup dialog box will open: On the first screen, select the COM port, connection method and address of the desired data logger. […]
4100-iSIC
iSIC Maintenance
Internal Battery Reorder Part Number: A01 Recommended Maintenance Frequency: 36 Months (min.) or As Needed The NexSens iSIC data logger comes standard with a 12V, 8.5 A-hr rechargeable sealed lead acid (SLA) battery. This battery will need to be replaced every 24-36 months. Any battery older than 36 months should be replaced immediately. This ensures […]
Antennas
High gain antennas are used for all deployments with iSIC data loggers using telemetry. The antennas are mounted on the iSIC box. If the antenna is to be mounted on a pole, the mounting pole should have a diameter up to 2-1/2 inches and have a height as high as practical for the application; also […]
Stainless Steel Enclosures
AVSS241608 Stainless Steel Enclosure The stainless steel enclosure (Part # AVSS241608) is ideal for NexSens iSIC Data Loggers installed in harsh environments or vandal-prone areas. This enclosure may also be used to house the NexSens 26 A-Hr battery (Part # A03) for projects with power-hungry sensors and heavy cloud coverage. Each enclosure also comes with […]
Powering the iSIC
The NexSens iSIC is powered from an internal battery. The addition of a float charger or solar panel can provide constant re-charging of the battery for prolonged use scenarios. Before data can be retrieved from sensors, power must be supplied to the iSIC. To do so, open the iSIC enclosure and unscrew the battery mounting […]
Grounding Protection
Electronic circuits are susceptible to voltage surges. The iSIC has on-board transient voltage suppression devices. Connect the ground lug to a proper electrical ground (i.e. earth ground or a properly grounded pipe/metal fixture). For outdoor applications, use the copper wire supplied with the A38 ground kit to connect the ground lug to the rod’s ground […]
Mounting the iSIC
NexSens recommends the A55 mount kit be used for all outdoor applications when mounting to metal poles, wooden posts, or walls. The A55 kit includes a ¼” aluminum plate, u-bolts, and lag screws. Use the supplied u-bolts for mounting to metal poles less than 2 3/8” diameter. Attach the plate to the enclosure using the […]
Monitoring Critical Projects
Due to the nature of many environmental projects, immediate notification of failures is a necessity. The iChart System Monitor allows a project manager to receive critical messages such as failure to retrieve data or upload data online. Click Project | Project Monitor. The System Monitor will appear. The Status tab of the System […]
iSIC Alarms
iSIC alarms are used to change the functionality of the data logger based on parameter inputs, such as changing sample and log intervals based on a particular logged reading. Open iChart with the current project. Click Setup. In the Setup Device window, click the iSIC Alarm tab. Place a check in the […]
iSIC Control Outputs
Parameter alarms configured on the iSIC data logger allow users to setup and configure controls to do things such as flash warning lights, trigger ISCO samplers, etc. The iSIC allows two kinds of control, 3.3V digital I/O and 12V 100mA switches. These can be accessed via the Setup Device Wizard in iChart. This may be […]