YSI DCP Adapter Wiring for Connection to X2 Data Logger
This article serves as a guide for wiring in a YSI EXO 1 or 2 sonde into a YSI DCP Adapter for connection to an X2 data logger. This configuration typically occurs within a NexSens A50 Junction Box to protect the DCP adapter and flying lead cables from environmental disturbances.
YSI DCP Adapter Wiring
NexSens UW8-FLx Cable* | DCP Adapter Connection | YSI EXO flying lead field cable** | NexSens MCIL-6-FS flying lead cable** | ||||
UW Plug Pin | Wire Color | Left Terminal | Right Terminal | Wire color | Male/Female Impulse Pin | Wire Color | Female Impulse Pin |
6 | Brown | SDI-12 | — | — | — | — | — |
2 | Black | GND | BLK | Black | 2 | Black | 2 |
5 | Red | 9-16VDC IN | RED | Red | 3 | Red | 3 |
— | — | — | BARE | Bare | 4 | Brown | 4 |
— | — | — | YEL | Yellow | 5 | Yellow | 5 |
— | — | — | WHT | White | 6 | White | 6 |
— | — | — | — | — | — | — | — |
— | — | — | — | — | — | — | — |
*Connection to left terminal only on DCP adapter.
**Connection to right terminal only on DCP adapter.
|
- YSI EXO Field Cable
- YSI EXO Flying Lead Field Cable
- NexSens MCIL Cable Assemblies
- NexSens UW8 plug to Flying Lead