YSI DCP Adapter Wiring for Connection to X2 Data Logger


This article serves as a guide for wiring in a YSI EXO 1 or 2 sonde into a YSI DCP Adapter for connection to an X2 data logger. This configuration typically occurs within a NexSens A50 Junction Box to protect the DCP adapter and flying lead cables from environmental disturbances.

 

YSI DCP Adapter Wiring

 

NexSens UW8-FLx Cable*DCP Adapter ConnectionYSI EXO flying lead field cable**NexSens MCIL-6-FS flying lead  cable**
UW Plug PinWire ColorLeft TerminalRight TerminalWire colorMale/Female Impulse PinWire ColorFemale Impulse Pin
6BrownSDI-12
2BlackGNDBLKBlack2Black2
5Red9-16VDC INREDRed3Red3
BAREBare4Brown4
YELYellow5Yellow5
WHTWhite6White6

*Connection to left terminal only on DCP adapter.
**Connection to right terminal only on DCP adapter.

 

Pin numbering for
NexSens UW8 Plug.

 

Pin numbering for NexSens MCIL-6-FS cable. Pin outline matches both NexSens and YSI field cables.

Pin numbering for MCIL-6 Male Connector from EXO field cable. Pin outline matches both NexSens and YSI field cables.